| KALTGAS
is a tempering system that utilizes the very low temperature of
liquid nitrogen as a cooling agent. The jet vaporizes the liquid
nitrogen in a cryogenic container, producing a steady cryogenic
gas flow. The LN2 vaporizer jet can be adjusted to minimize liquid
nitrogen consumption and to adjust both the cooling rate and the
gas flow.The cryogenic gas flow is then piped through a vacuum
insulated flexible metal line to the object to be cooled. |
 |
A
KALTGAS system needs only a few minutes to produce a cold gas flow
with a temperature as low as -180°C (-292°F). In most
cases, this KALTGAS system is only used to produce a cold gas flow.
The volume of the gas flow is adjusted with the safety control
unit. Since the controlled system operates without a heat exchanger,
it is not possible to maintain precise temperature stability.
To achieve good temperature and control stability,
a post-heating module or heat exchanger is installed. This module
can be connected directly to the line, or installed in a chamber
adjacent to the cold gas flow. The heat exchanger is designed
to heat the cold gas flow to the desired temperature. The final
product, a clearly quantified gas flow exiting the heat exchanger
at a defined temperature, is ready for use as a cooling agent.
With an optimized temperature controller a temperature stability
of ±0.1°C can be achieved. With the standard safety
control unit, a temperature stability of better than ±0.2°C
can be achieved at a steady state.
Apart from the high cooling speed, another
advantage of KALTGAS systems is their modular design. By swapping
out individual modules such as the N2 gas line, the
LN2 vaporizer or the heater, it is possible to change
the cooling speed, LN2 consumption as well as the
application. The basic modules, including the LN2 container,
the vacuum pump and the safety control unit, remain unchanged. |
| TG50
Kaltgas Cryogenic System, Vacuum Pump, SC2 Control Unit Cryogenic
cooling system for small cooling applications to -170°C
(-335° F) The TG 50 KALTGAS system includes a temperature
controller with current value/set point display, a safety controller,
a KF-NW 50 siphon with an LN2 vaporizer (jet), a
flexible, vacuum insulated N2 gas line, a PT100
temperature sensor and a vacuum pump with accessories. |
| TG50-FV Kaltgas Cryogenic System, Evacuated Tube, SC2 Control Unit Cryogenic
cooling system for small cooling applications to -170°C
(-335° F) The TG 50 KALTGAS system includes a temperature
controller with current value/set point display, a safety controller,
a KF-NW 50 siphon with an LN2 vaporizer (jet), a
flexible, vacuum insulated N2 gas line, a PT100
temperature sensor. No vacuum
pump. |
| TG-DD
Kaltgas Cryogenic System with Vacuum Pump Cryogenic
cooling system for continuous cooling applications from +120°C
to -160°C This KALTGAS system is designed for continuous
cooling in a temperature range from -160°C to +120°C,
in applications where a gas flow is directed at the sample
to be cooled. |
| TGM10 Kaltgas
Cryogenic System, Vacuum Insulated Tube 5 liter LN2 Container For
cryogenic cooling of small specimens. The TGM KALTGAS system
includes an output controller with current value display, a
23AL-NF60 Dewar flask, an NW 60 siphon head with an LN2vaporizer,
a vacuum insulated N2 glass tube with swivel
and protective cover, and an integrated Ni Cr Ni temperature
sensor. |
| TGL63-50
Kaltgas Cryogenic System, Vacuum Pump,SC2 Control Unit The
TGL63-50 KALTGAS system includes a temperature
controller with current value/set point display, a safety
controller, a KF-NW 50 siphon with an LN2 vaporizer
(jet), a flexible, evacuateable N2 gas line
with an integrated post heating module (heater), a PT100
temperature sensor and a vacuum pump with accessories. |
| TG63-100
Kaltgas
Cryogenic System, Vacuum Pump,SC2 Control Unit The
TGL63-100 KALTGAS system includes a temperature controller
with current value/set point display, a safety controller,
a KF-NW 50 siphon with an LN2 vaporizer (jet), a
flexible, evacuateable N2 gas line with an integrated
post heating module (heater), a PT100 temperature sensor and
a vacuum pump with accessories. |
| TGL63-50KF
Kaltgas Cryogenic System, Vacuum Pump,SC2 Control Unit The
TGL63-50KF KALTGAS system includes a temperature controller
with current value/set point display, a safety controller,
a KF-NW 50 siphon with an LN2 vaporizer (jet),
a flexible, evacuateable N2 gas line with an integrated
post heating module (heater), a PT100 temperature sensor
and a vacuum pump with accessories |
| TGL63-100KF
Kaltgas
Cryogenic System, Vacuum Pump,SC2 Control Unit The
TGL63-100KF KALTGAS system includes a temperature controller
with current value/set point display, a safety controller,
a KF-NW 50 siphon with an LN2 vaporizer (jet), a
flexible, evacuateable N2 gas line with an integrated
post heating module (heater), a PT100 temperature sensor and
a vacuum pump with accessories |
| TGL63-50KKK
Kaltgas Cryogenic System, Vacuum Pump,SC2 Control Unit The
TGL63-50KKK KALTGAS system includes a temperature controller
with current value/set point display, a safety controller,
a KF-NW 50 siphon with an LN2 vaporizer (jet), a
flexible, evacuateable N2 gas line with an integrated
post heating module (heater), a PT100 temperature sensor and
a vacuum pump with accessories |
| TGHD40-50
Kaltgas Cryogenic System, Vacuum Pump,SC2 Control Unit The
TGDH 40/50 KALTGAS system includes a temperature controller
with current value/set point display, a safety controller,
a KF-NW 50 siphon with an LN2 vaporizer
(jet), a flexible, evacuateable N2 gas line with
an integrated post heating
module (heater), a PT100 temperature sensor and a vacuum pump
with accessories. |
| TGDD
Kaltgas System Vacuum Pump The TGDD KALTGAS
system includes an vaporizer Dewar with an LN2 vaporizer
(jet), an N2 post heating unit, an integrated
min/max sensor for LN2 refilling, a
temperature controller with current value/set point display,
a safety controller, an evacuateable N2 gas line
with an integrated post heating module
(heater, low voltage design), an integrated PT100 temperature
sensor, a vacuum pump with accessories, and a conventionally
insulatedLN2 line withLN2 tank
siphon and LN2 solenoid valve. |
| TGRD40-50
Kaltgas Cryogenic System, 2 liter Reaction Vessel, Frame, Lid In
a reaction vessel, 1.5 liters of methanol are cooled from
+20°C to -95°C in approx. 50 minutes. During
the cooling phase, about 5 to 6 liters of LN2 are
consumed. To maintain the methanol at -95°C, the reactor
needs only about 1.2 liters of LN2 per hour. |
| TGRID40-50
Kaltgas Cryogenic System, 2 liter Reaction Vessel, Frame, Lid,
SC2 Control Unit In
a reaction vessel, 1.5 liters of methanol are cooled from
+20°C to -95°C in approx. 50 minutes. During the
cooling phase, about 5 to 6 liters of LN2 are
consumed. To maintain the methanol at -95°C, the reactor
needs only about 1.2 liters of LN2 per hour. |
| SC2
Single Channel Control Unit |
| GR1
Gas Regulator Unit |
|
|